Important Safety Information
Regarding Y
our Insulin Pump
(includes all models)
Avoid Immersing Your Pump In Water
The pump was tested and met requirements for IPX7 at time of manufacture. For more information about
IPX7, see the Glossary on page 229. Be aware that drops and bumps that occur over time will affect the
integrity of the pump case and make it more vulnerable to damage from water. Lotions, sunscreens and
insect repellent can also damage the pump case. It is not known how much water is needed to damage
your pump. You should avoid getting your pump wet. To shower, bathe, swim, or participate in water
activities, always disconnect from your pump and reconnect after you are out of the water.
If you inadvertently submerge your pump in water, dry the pump quickly using a soft, clean towel and
verify that it is working properly by selecting Selftest from the pump's Utilities Menu. If you believe that
water has entered your pump or you observe any other possible pump malfunction, please check your
blood glucose, treat high blood glucose (if necessary) with an injection and contact our 24 Hour HelpLine
at 1-800-646-4633 for further assistance. Symptoms of high blood glucose include fatigue, excessive thirst
and nausea. Always contact your healthcare professional if you experience excessively high or low blood
glucose levels, or if you have any questions about your care.
Electrostatic Discharge
Although your Paradigm pump is designed to be unaffected by typical levels of electrostatic discharge
(ESD), very high levels of ESD can result in a reset of the pump's software with an associated pump error
alarm. In most cases, exposure to high levels of ESD will trigger the pump's A-13 alarm although, under
certain circumstances, high level ESD exposure can cause A-44, Bolus Stopped or Max Delivery alarms. High
levels of ESD are more likely in situations where the relative humidity is very low, such as inside a heated
building during the winter in areas where it is cold outside.
If your pump experiences an A-13 or other error alarm, press the ESC and ACT buttons to clear the alarm.
If you are unable to clear the alarm by pressing ESC and ACT, you may need to remove and replace the
pump's battery to clear the alarm. After clearing the alarm, you should always verify that your pump is set
to the correct date and time and that all other settings (basal rate, max basal and bolus limits, etc.) are
programmed to the desired values, since the software reset could erase your previously programmed
settings. Please see the Alarms and Alerts section of this User Guide for more details regarding what to do
if your pump displays an error alarm or other alert message.
Please contact our 24 Hour HelpLine at 1-800-646-4633 to report any error alarms or other problems that
occur with your pump.