30°
Step 5
PLACE the feet of the
Sil-Serter on the insertion
site and position at an
approximately 30 degree angle.
Step 7
OPEN the Sil-Serter mouth
again and SLIDE it backwards.
Step 10
CONNECT the tubing and the
Silhouette set.
Recommended Cannula Fill:
0.7 U for 13 mm and 17 mm
Step 6
INSERT the Silhouette set by
pressing the white button.
Step 8
WITHDRAW the introducer
needle by squeezing on the
sides before pulling.
Step 9
REMOVE the other half of
protective lm from the part
before the adhesive.
ADHERE the back part of the
adhesive to the skin.
Refer to Chapter 5 in the MiniMed Paradigm™ Veo™ User Guide
(Insertion of the Medtronic Silhouette Infusion Set, continued)