Medtronic Guardian Real-time Guia do Utilizador

Consulte online ou descarregue Guia do Utilizador para Medição, teste e controlo Medtronic Guardian Real-time. Medtronic Guardian Real-time User guide Manual do Utilizador

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir

Resumo do Conteúdo

Página 1 - User Guide

CONTINUOUS GLUCOSE MONITORING SYSTEMUser Guide

Página 2 - REAL-Time

AUC calculation ... 31Setting the Alarm Snooze ...

Página 3 - 6025254-019_c

LOST SENSORReason: The monitor has not received a signal from the transmitter for more than 40 minutes.Action: Make sure that the sensor is inserted

Página 4

Action: If the alert resulted from two Cal Errors in a row, replace the sensor. If the alert happenedwithout two Cal Errors, use the transmitter'

Página 5

METER BG BYReason: A meter BG measurement is required for calibration purposes by the time that is shown on yourmonitor. The METER BG BY alert is al

Página 6

FALL RATEReason: The sensor glucose measurements are falling at a rate that is equal to or faster than the Set FallRate Limit you selected for the a

Página 7

Sensor Alarm History display = this alert• CalErr = CAL ERROR• ChgSen = CHANGE SENSOR• ChgXtr = CHARGE TRANSMITTER•DwnSlp = FALL RATE•Hi = HIGH SG•Lo

Página 8 - Contents

System alarmsAlarms put the monitor in Attention mode. You may encounter the alarms listed below while using thesystem. These alarms include informati

Página 9 - Contentsviii

FAILED BATT TESTThe battery does not have enough energy to make the monitor work and the battery cannot be used.Install another battery, and always

Página 10 - Contents ix

• BadBat = FAILED BATT TEST•ChkSet = CHECK SETTINGS• Exx = Any E-Alarm• NoPwr = OFF NO POWER• Reset = RESET• WeakBt = WEAK BATTERY To view your system

Página 11 - Contentsx

Question: Why do I see a mix of quick andlong flashing red lights on the charger?Answer: Your charger and transmitterbatteries are very low. Replace t

Página 12 - Contents xi

Connecting the tester1 Hold the transmitter and the tester as shown. Line up the flat side of the testerwith the flat side of the transmitter.2 Push t

Página 13 - Contentsxii

3 hour graph ... 576 hour graph ...

Página 14 - Contents xiii

Troubleshooting the monitorMy monitor will not display my BG measurement from my meter1 Make sure you are using the correct meter (a blood glucose met

Página 15 - Contentsxiv

MAIN MENU > Sensor > Sensor StartThe SENSOR START MENU appears with New Sensor selected.2 Select Find Lost Sensor.Press ACT. A message tells you

Página 16 - Introduction

Why doesn't my monitor battery last very long?• A short battery life does not always mean something is wrong with your monitor. Battery life in y

Página 17 - Your system

I submerged my monitor in waterYour monitor is designed to resist brief, accidental contact with water. Do not submerge the monitor inwater when bathi

Página 18 - Accessories

System maintenanceReplacing the batteryYou must replace the monitor battery with a new battery as part of your normal system maintenance. Seethe Insta

Página 19 - Introduction4

Cleaning the systemThis procedure shows you how to clean the different parts of your system. To avoid accidentally damagingyour system during cleaning

Página 20 - User safety

CAUTION:• Never place your monitor directly in water.• Keep the battery compartment dry and away from moisture.• Never try to clean the inside of your

Página 21 - Safety Warnings

For details on attaching and disconnecting the tester, see Connecting the tester, on page 94 andDisconnecting the tester, on page 94.Inspecting the tr

Página 22 - Adverse reactions

For single-patient useAlways clean the transmitter after each use.To clean the transmitter, you will need the following materials: Ivory® liquid soap,

Página 23 - Compliance information

5 Prepare a mild liquid soap solution using one (1) teaspoon of Ivory liquid soap per one (1) gallon ofroom temperature tap water.6 With tester still

Página 24 - Introduction 9

Logbook ... 82Chapter 9: Troubleshooting and alarms/

Página 25 - Introduction10

10 Holding the tester, wipe the transmitter with 70% isopropyl alcohol.11Place the transmitter and tester on a clean, dry cloth and air dry them compl

Página 26 - The basics

For multiple-patient useWhen using the transmitter in a healthcare facility, always clean and disinfect the transmitter after eachuse.Warning: You mus

Página 27 - Battery cap

Warning: If you see any body fluid in the connector opening, you must discard the transmitter.Because the transmitter contains a battery, do not disca

Página 28 - The monitor buttons

8 Holding the tester, remove the transmitter from the solution. Brush the entire surface of thetransmitter using a soft-bristled brush, paying close a

Página 29

13 Rinse the transmitter under room temperature tap water for three minutes.14 Holding the tester, wipe the transmitter with 70% isopropyl alcohol.15

Página 30 - The monitor screen

18 Inspect the housing of the transmitter for any signs of cracking, flaking, or damage. If you see any ofthese signs, you must now discard the disinf

Página 31 - Screen backlight

System specificationsAlarm/alert types• Audible tone (beep) — To make sure that you are aware of the alarm/alert, the sound goes from abeep to a siren

Página 32 - BATT TEST

Sensor glucose graph timeout• Default: 2 minutes•Options•2 minutes•4 minutes•6 minutes• NONE (no timeout)High and low Glucose Limits (CSS7100)• Maximu

Página 33 - UTILITIES MENU

Daily totals• Maximum number of daily totals recorded: 31 daysDefault screen• The HOME screen. The time always shows across the top of the HOME screen

Página 34 - The STATUS screen

Feature Options Default Setting Limits, Range of ValuesAUC Limits Always on Low Limit 70 (mg/dL) 3.9 (mmol/L)High Limit 180 (mg/dL) 10.0 (mmol/L)Low L

Página 35 - The basics20

Connecting the tester ... 94Disconnecting the tester ...

Página 36 - Entering your settings

Feature Options Default Setting Limits, Range of ValuesGlucose Alarms•On•OffOff You set the thresholds(limits) for the GlucoseAlarms using the Glucose

Página 37 - Month Day

Feature Options Default Setting Limits, Range of ValuesGlucose Limits2throughGlucose Limits 8CSS7100Low: Off, OnHigh: Off, OnCSS7100KLow• Off, or vali

Página 38 - Selecting the language

Feature Options Default Setting Limits, Range of ValuesLanguage•Cesky•Dansk•Deutsch•English• Español•Français•Italiano•Magyar• Nederlands• Norsk•Polsk

Página 39 - Setting your Alarm/Alert Type

Feature Options Default Setting Limits, Range of ValuesMissed Data (Always on - cannot beturned off)30 minutes Range:0:05 – 0:40 (5 to 40 minutes)Pred

Página 40 - Turning on the Glucose Alerts

Feature Options Default Setting Limits, Range of ValuesSensor glucosegraph timeout• 2 minutes• 4 minutes• 6 minutes•NONEa2 minutes Possible values: Sh

Página 41 - Setting the Glucose Limits

Radio frequency (RF) communication specifications• Monitor to transmitter frequency: 916.5 MHz; proprietary Medtronic protocol; range up to 6 feet(1.8

Página 42 - Entering your settings 27

Monitor size• Inches•Heightapproximately 2.0 inches•Widthapproximately 3.2 inches• Depth (thickness)approximately 0.77 inches• Centimeters (cm)•Height

Página 43 - Setting the High Snooze

Screen item DescriptionBattery: Shows battery condition (how much energy is left in the battery). This alwaysappears.BG meter value: Shows the most re

Página 44 - Setting the Low Snooze

Guidance and manufacturer's declarationGuidance and Manufacturer's Declaration - Electromagnetic EmissionsThe Guardian REAL-Time system (con

Página 45 - Setting the Rate Alerts

Guidance and Manufacturer's Declaration - Electromagnetic ImmunityThe Guardian REAL-Time system (consisting of the CSS7100/CSS7100K Receiver and

Página 46 - AUC calculation

High and low Glucose Limits (CSS7100) ... 112High and low Glucose Limits (CSS7100K) ...

Página 47 - Entering your settings32

Guidance and Manufacturer's Declaration - Electromagnetic ImmunityThe Guardian REAL-Time system (consisting of the CSS7100/CSS7100K Receiver and

Página 48 - Setting the Cal Reminder

Guidance and Manufacturer's Declaration - Electromagnetic ImmunityThe Guardian REAL-Time is intended for use in the electromagnetic environment s

Página 49 - SN GT2XXXXXXM

Guidance and Manufacturer's Declaration - Electromagnetic ImmunityThe Guardian REAL-Time is intended for use in the electromagnetic environment s

Página 50 - Reviewing your settings

Recommended separation distances between portable and mobile RF communicationsequipment and the Guardian REAL-Time systemThis section provides informa

Página 51 - Using the Sensor Demo

MiniLink specificationsBiocompatibility MiniLink: Complies with ISO 10993-1 for long-term body contactOperating Conditions MiniLink Temperature: +32°

Página 52 - Transmitter charger

WarrantyMedtronic MiniMed warrants the Medtronic MiniMed Guardian® REAL-Time Continuous Glucose Monitoring(CGM) System against defects in material and

Página 53 - Charging the transmitter

Icon table Attention: Read all warnings and precautions in instructions for use.Date of manufacture (year - month)ManufacturerBatch codeUse by (year -

Página 54 - Inserting the sensor

Magnetic Resonance (MR) unsafe:System specifications132

Página 55 - Starting the sensor

Sensor accuracyThe information in this section is meant to be reviewed by you with your healthcare professional. Pleasealso note: The Guardian RT is n

Página 56

Performance results in adultsThe performance of the Guardian RT in adults was evaluated in a clinical study.3 Guardian RT results werecompared to plas

Página 57

Glossary ... 155Index ...

Página 58 - Meter option

Agreement between the matched pair was estimated by evaluating the difference between the GuardianRT reading and the YSI measurement. The difference b

Página 59 - Add, delete, review meter IDs

The Clarke Error Grid was used to assess the clinical relevance of the differences between the Guardian RTreadings and the comparative YSI measurement

Página 60 - Meter option 45

The plot below is a correlation plot of Guardian RT readings versus readings from the reference method,the YSI 2300 Glucose Analyzer. It is overlaid w

Página 61 - Meter option46

The percent of Guardian RT readings in the previous graph are presented in the following table accordingto the percentage of points falling within eac

Página 62 - Using your system

Time of dayGlucose Conc (mg/dL)The above chart, Guardian RT device #2 is represented by a solid line, and the Guardian RT device #4 by alight-gray lin

Página 63 - Using your system48

The Low Glucose AlertThe Low Glucose Alert was evaluated for its ability to detect glucose levels at 70 mg/dL (3.9 mmol/L), orbelow, using the YSI 230

Página 64 - Using your system 49

high blood glucose can be obtained by setting the High Glucose Alert level lower. For example, setting theHigh Glucose Alert at 190 mg/dL (10.6 mmol/L

Página 65 - Capturing Events

As per the stratified Clarke Error Grid analysis above, agreement between Guardian RT values and YSIvalues tends to be less at low and high glucose co

Página 66 - Using your system 51

An analysis of the mean percentage of Absolute Relative Error (ARE percentage) and standard deviations,across 12-hour increments of wear periods, appe

Página 67 - Before you begin

The percentage of Guardian RT readings within 20 percent and 30 percent of YSI readings was analyzedaccording to time after sensor insertion and accor

Página 68 - Entering exercise information

IntroductionThank you for choosing Medtronic Diabetes (legally known as Medtronic MiniMed, Inc.) as your diabetesmanagement partner.Your new Guardian®

Página 69 - Entering Other markers

Time-Elapsed PlotsThe plot graph below is a representative example of continuous sensor tracing vs. reference blood glucosereading, where sensor showe

Página 70

The plot graph below is a representative example of continuous sensor tracing vs. reference blood glucosereading, where sensor showed average (typical

Página 71 - The graphs

The plot graph below is a representative example of continuous sensor tracing vs. reference blood glucosereading, where sensor showed poor performance

Página 72 - 3 hour graph

Agreement between the matched pair was estimated by evaluating the difference between the GuardianRT reading and the blood glucose meter measurement.

Página 73 - 6 hour graph

The accuracy of the Guardian RT was also evaluated by calculating the percentage of Guardian RTreadings within 20% and within 30% of the blood glucose

Página 74 - 12 hour graph

The following figure is a correlation plot of Guardian RT readings versus readings from the referencemethod, the blood glucose meter. It is overlaid w

Página 75 - 24 hour graph

Stratified Clarke Error Grid AnalysisRange ofComparativeGlucoseReadings(mg/dL)TotalCount A + B A B C D E40-80 360 (13.9%) 201(55.8%)141(39.2%)60 (16.7

Página 76 - The rapid change arrows

low blood glucose can be obtained by setting the Low Glucose Alert level higher. For example, setting theLow Glucose Alert at 90 mg/dL (5.0 mmol/L), i

Página 77 - Checking the sensor status

Sometimes the Guardian RT will alert when the blood glucose levels are not high. When the Guardian RTHigh Alert was set at 250 mg/dL (13.8 mmol/L) in

Página 79 - Using your system64

Department Telephone number24 Hour HelpLine (calls within the United States) 800.646.463324 Hour HelpLine (calls outside the United States) +1 818 576

Página 80 - Using your system 65

Glossary AAdverse reaction - Anunexpected, unpleasant ordangerous reaction to asensor when it is inserted intothe body. An adverse reactionmay be sudd

Página 81 - Viewing your current settings

METER BG BY alert will occurtwo hours before thecalibration is required.Calibrate - Check, adjust, or setto a standard (calibrate yourmonitor).Circums

Página 82 - Using your system in water

High Snooze - For the HighGlucose, High Predictive, andRising Rate of change alerts,allows you to set the delaybetween the first alert andany subseque

Página 83 - Using your system68

PPredictive alerts - These alertscan be used so that youreceive alarms before thesensor glucose measurementsreach your glucose limits. Thisalert predi

Página 84 - Utilities

Sterile gauze - A thin medicalbandage with loose weavethat is free from substancesor germs that could causeinfection.Submerge - To place underwater or

Página 86 - Utilities 71

Index Numerics3 hour graph 576 hour graph 5812 hour graph 5924 hour graph 60AA (alarm) condition 90accessoriesCareLink USB 3holster 3leather ca

Página 87 - The Alarm Clock

Bbacklight 13, 14, 16, 88, 97specifications 111bad transmitter battery 92Batt Out Limit alarm 11Batt out limit alarmcondition 90battery 16, 17,

Página 88 - Utilities 73

Glucose Alarms 115Glucose Limits 115Glucose Limits 1 115Glucose Limits 2 through8 116High Glucose Limit 116High Snooze 116Language 117Lock Keyp

Página 89 - The Block feature

High Predictive alertcondition 87High Snoozedefault settings 116setting 28High Snooze alarmsetting 28historysensor updates 65How to make glucose

Página 90 - Running the Selftest

• Sensor — The sensor (MMT-7002/MMT-7003) is a membrane-coveredelectrode inserted through the skin into the fatty layer underneathwith an insertion de

Página 91 - User settings

Next Cal 63Normal mode 17OOff no power alarm condition 91on an aircraftentering meter bloodglucose measurements 8OTHER screen 54PPrecautions 7ex

Página 92 - Restoring your settings

Sensor icon 15Sensor Isig 63SENSOR MENU 18SENSOR ON/OFF screen 25, 36SENSOR READY 2 HRS screen 40SENSOR STATS screen 71, 72sensor status 62Next

Página 93 - Clearing your settings

Storing the system 99supplies, ordering 3System alarm conditions 90system alarm/alert historyviewing 91system settingsclearing 76, 78restoring 7

Página 96 - Therapy management software

Conventions used in this guideThe conventions used in this guide are defined in the following table.Conventions MeaningPress Push and release the butt

Página 97 - Therapy management software82

Guardian® REAL-TimeContinuous Glucose Monitoring SystemUser Guide

Página 98 - About alarms and alerts

User safetyThis section includes important safety information such as indications, contraindications, safety warnings,potential adverse reactions, and

Página 99 - Sensor alerts

The Guardian REAL-Time CGM System (CSS7100K) provides real-time glucose values that allow users totrack patterns in glucose concentrations and to poss

Página 100 - CHANGE SENSOR

MeterDo not use an RF-capable meter to send blood glucose measurements to your monitor while traveling on anaircraft or if it interferes with another

Página 101 - METER BG NOW

Damage due to radiationIf you are going to have an X-ray, MRI, diathermy treatment, CT scan, or be exposed to other types ofradiation, REMOVE YOUR MON

Página 102 - HIGH PREDICTED

These devices generate, use, and can radiate radio frequency energy and may cause interference to radioor television communications, even if installed

Página 103 - LOW BATTERY

RF interference from other devicesCommon consumer electronic devices that transmit in the same frequency band used by the MMT-7703MiniLink transmitter

Página 104

The basicsCAUTION: Never use sharp objects to press the buttons on your monitor as this can damage thebuttons or compromise the seal on the monitor. S

Página 105 - System alarms

Battery life varies widely depending on the way you use your monitor. Some of the use factors that affectbattery life include:• the alert type you cho

Página 106 - WEAK BATTERY

6 If the HOME screen appears, then go to the next step. If the HOME screen does not appear:a. Make sure that the battery is inserted correctly. If not

Página 107

Button Description• Accept the selected menu item.• Choose a setting when entering or changing settings.• Open the MAIN MENU from the HOME screen.• Cl

Página 108

©2006, Medtronic MiniMed, Inc. All rights reserved.This product is covered by U.S. Patents. Other U.S. and/or foreign patents may be pending. Pleasevi

Página 109 - Disconnecting the tester

The monitor screenThe screen shows five lines of text at one time. The first is the operating mode. The second is the currentopen menu or function. Th

Página 110 - Troubleshooting the monitor

Battery iconThis icon shows when a battery is installed in the monitor. Thereare four segments in the icon. Each segment representsapproximately 25 pe

Página 111

Alarm/alert typesYou can set up your monitor to either beep or vibrate when an alarm or an alert occurs. See the nextchapter for details on setting th

Página 112 - I dropped my monitor

The screen shows the name of the alarm and information about how to clear it (press to see theinformation on clearing the alarm). In this example, t

Página 113

The STATUS screenThe STATUS screen shows any recent activity, the condition of the monitor battery, and which specialfeatures are active, if any.• To

Página 115 - Cleaning the system

Entering your settingsBasic stepsPerform these basic steps in the order shown.1 Make sure that you have installed the battery in your monitor.2 Make s

Página 116 - Watertight tester

4Use / to set the correct time and date as follows:a. Set the correct hour, then press ACT.b. Set the correct minutes, then press ACT.c. Set the co

Página 117 - System maintenance102

Selecting the languageIf you do not read English, the next step is to change the Language setting so that all text information,alarms, alerts, sensor

Página 118 - For single-patient use

Setting your Alarm/Alert TypeYou can set your monitor three different ways to let you know when there is an alarm, alert, or specialconditions or prog

Página 119 - System maintenance104

Contacts: Africa:Medtronic Africa (Pty) Ltd.Tel: +27 (0) 11 677 4800Australia:Medtronic Australasia Pty. Ltd.Tel: 1800 668 670 (product orders)Tel: 18

Página 120 - System maintenance 105

Turning on the sensorThe sensor must be turned on, started and initialized to report glucose measurements.1 Open the EDIT SETTINGS screen.MAIN MENU &g

Página 121 - For multiple-patient use

Setting the Glucose LimitsAfter you turn the Glucose alerts On, then you need to set the high and low Glucose Limits recommendedby your healthcare pro

Página 122 - System maintenance 107

12:00 AM 9:00 AM12:00 AMGlucose Limits 1High Glucose Limit: 240 mg/dL (13.3 mmol/L)Low Glucose Limit: 60 mg/dL (3.3 mmol/L)12:00 AM - 9:00 AM(9 hour

Página 123 - System maintenance108

10 Select your Low Glucose Limit. The available range is:• 40 to 390 mg/dL (2.2 to 21.6 mmol/L) for CSS7100• 90 to 390 mg/dL (5.0 to 21.6 mmol/L) for

Página 124 - System maintenance 109

4 Press ACT. The EDIT SETTINGS screen shows that the High Snooze is set.Setting the Low SnoozeAfter you receive and clear a Low Glucose, Fall Rate of

Página 125 - Removing adhesive residue

4 Select the Time Sensitivity (how many minutes before you reach your Low Glucose Limit).•Range:5 to 30 minutes (0:05 to 0:30)• Increments:5 minutes5

Página 126 - System specifications

3 Select the Fall Rate Limit.•Range:mmol/L/minmg/dL/minfrom 0.061 to 0.277from 1.1 to 5.0• Increments:mmol/L/minmg/dL/min0.0050.14 Press ACT. The SET

Página 127 - Sensor glucose graph timeout

12:00 A 4:00 A8:00 A 12:00 P4:00 P 8:00 P12:00 A4003502802101407040mg/dL (mmol/L)AUC High Average = 30 mg/dL (1.7 mmol/L) AUC Low Average = 3.9 mg/

Página 128 - Default settings

Setting the Alarm SnoozeAfter you receive and clear a METER BG NOW alert, the monitor will repeat the alert until you enter a newmeter blood glucose m

Página 129

Entering the transmitter identification numberThe Transmitter ID starts with 2 and is found on the flat side of your transmitter. You must enter thetr

Página 130

Latvija:Ravemma Ltd.Tel: +371 7273780Macedonia:Alkaloid Kons DooelTel: +389 2 3204 430Magyarország:Medtronic Hungária Kft.Tel: +36 1 889 0688Malaysia:

Página 131

4 Press ACT. The EDIT SETTINGS screen shows the length of time you entered for the Missed Dataoption.Setting the sensor glucose graph timeoutThe senso

Página 132

4 Save your settings after you have set your preferences so that you can restore the settings you havesaved if you receive an alarm/alert or error tha

Página 133

Setting up your transmitter and sensorBasic stepsAfter you have set up your monitor, there are a few basic steps that you must complete to set up your

Página 134 - System safety checks

2 Insert a new AAA or LR-03 battery with the flat (-) end first. Make sure that youalign the small bumps on the battery cap with the small notches in

Página 135 - Data security

green light4 When charging is complete, the green light on the charger will stay on, without flashing, for 15 to20 seconds and then turn off.5 After t

Página 136 - STATUS screen

3 Hold the transmitter as shown. Line up the two notches on the transmitter withthe side arms of the sensor. The flat side of the transmitter should f

Página 137 - System specifications122

Entering the first meter blood glucoseYou cannot enter a meter blood glucose (meter BG) measurement into your monitor until you receive aMETER BG NOW

Página 138 - System specifications 123

Setting up your transmitter and sensor42

Página 139 - System specifications124

Meter optionAbout metersBayer's CONTOUR® NEXT LINK Wireless Meter (Model No. 6201/9624), the LifeScan OneTouch® UltraLink®meter, or the BD Paradi

Página 140 - System specifications 125

Meter rules If you want your monitor to communicate with the meter, the following conditions must apply:1 The meter option must be turned on and progr

Página 141 - System specifications126

Österreich:Medtronic Österreich GmbHTel: +43 (0) 1 240 44-024 – Stunden – Hotline: 0820 820 190Česká republika:Medtronic Czechia s.r.o.Tel: +420 233 0

Página 142 - System specifications 127

Reviewing meter IDsa. Select Review ID, then press ACT.b. The programmed IDs will show in the REVIEW METER ID screen.4Exit the menus when you are done

Página 144 - MiniLink specifications

Using your systemCalibrating your systemCalibrating your system helps to make sure that the glucose readings from the sensor are as accurate andreliab

Página 145 - Warranty

What are the best times to calibrate the system?It is suggested that you calibrate your system at the following times:•Before meals• Before going to b

Página 146 - Icon table

Time since last calibration What happens• The system will not take glucose measurements again until you enter anew Meter BG measurement. To calibrate

Página 147 - System specifications132

Capturing EventsThe Capture Event feature allows you to electronically save certain types of information that is used withMedtronic CareLink Personal

Página 148 - Sensor accuracy

4 Make sure that the number shown on the screen is correct.• If the information is correct, press ACT.• If the information is not correct, select No,

Página 149 - Performance results in adults

6 Make sure that the type and amount of insulin shown on the screen is correct.• If the information is correct, press ACT. The information you entered

Página 150 - Sensor accuracy 135

6 If the number on the screen is correct, press ACT. The information you entered is saved and can beused in reports. If the information is not correct

Página 151 - Appendix A136

Entering Other markersThis section shows you how to enter markers other than Enter BG, Insulin Marker, Meal Marker, or ExerciseMarker. When you enter

Página 153 - Appendix A138

• Reconnect old sensor (two hours).• A meter BG entry is needed for calibration.Two other conditions also prevent a real-time SG value and generate a

Página 154 - Time of day

The graphsAll graphs show High and Low Glucose Limits lines, a real-time SG measurement line, a data section, andthe cursor (flashing vertical line).

Página 155 - The High Glucose Alert

Examples of real-time sensor glucose graphsThe following pages show examples of the different graph types.3 hour graphEach part of the graph between t

Página 156 - Function of Time

6 hour graphEach part of the graph between the vertical dashed lines represents one hour. In this example, theselected SG measurement was taken at 1:1

Página 157 - Appendix A142

12 hour graphEach part of the graph between the vertical dashed lines represents one hour. In this example, LOSTSENSOR is the reason no measurement is

Página 158 - Sensor accuracy 143

24 hour graphEach part of the graph between the vertical dashed lines represents six hours. The shaded area helps youmore easily see the sensor glucos

Página 159 - Appendix A144

How to check for rapid changes in sensor glucoseIf your sensor glucose (SG) rises or falls faster than a certain rate, Rapid Change arrows automatical

Página 160 - Glucose Conc (mg/dL)

Checking the status of the systemThe system allows you to check its status (condition) in the following areas:• Condition of the batteries.• The age o

Página 161

Status DescriptionNext Cal Shows when your next calibration is due.Sensor Age The total number of days and hours that the sensor has been used.Sensor

Página 162

How to make glucose alerts silentThe Alert Silence feature allows you to make glucose alerts silent for a preset amount of time between30 minutes and

Página 163 - Appendix A148

ContentsChapter 1: Introduction ... 1Assistance ...

Página 164 - Sensor accuracy 149

5 Select the alert silence option you want, then press ACT. The SET DURATION screen flashes one of thefollowing:• A set of dashes (if this is the firs

Página 165

Viewing your current settingsYou can review your current settings at any time. This is useful for making sure that you use the correctsettings when tr

Página 166

Removing the sensorGently pull the sensor from your body to remove it. Place it in a sharps container.Using your system in waterYour monitor is NOT wa

Página 167

Using your system68

Página 168 - Sensor accuracy 153

UtilitiesThe data management featureThe system keeps a daily record of the following types of information for up to 31 days:• Daily totals — Daily tot

Página 169 - Appendix A154

Sensor glucose informationThe following table lists the different types of SG information that the system automatically records. TheSG information is

Página 170 - Glossary

Item DescriptionBG Average The average meter blood glucose measurement.Number BGs The total number of meter blood glucose measurements taken.Meter L/H

Página 171 - Glossary156

2 Press ACT. The DAYS TO AVERAGE screen flashes the number of days you selected the last time youused this feature.3 Select the number of days that yo

Página 172 - Glossary 157

Reviewing your current Alarm Clock remindersThis procedure shows you how to view your current Alarm Clock reminders. The dashes in this exampleshow th

Página 173 - Glossary158

The Block featureWhen this feature is turned on, anyone who may have access to your monitor cannot set or change any ofyour settings. The only feature

Página 174 - Glossary 159

The monitor screen ... 15The HOME screen ...

Página 175 - Glossary160

The Lock Keypad featureLock Keypad prevents accidental keypad presses. When the keypad is locked, you can still press ESC toview the STATUS screen, th

Página 176 - Numerics

3 Press ACT. The system automatically runs the Selftest in the following order:a. Screen Test:The screen appears all black.b. Selftest:The monitor cou

Página 177 - Index162

3 Press ACT. If you have saved your settings before, then a message shows the last time you saved yoursettings and asks if you want to overwrite them.

Página 178 - Index 163

Clearing your settingsUse this utility only if you want to clear your current monitor settings and restore the system defaultsettings.CAUTION: Do not

Página 179 - Index164

Viewing your User Settings historyThis feature allows you to view your most recent User Settings operations. The system can store up to 20User Setting

Página 181 - Index166

Therapy management softwareMedtronic provides software to optimize your monitor use. Contact our 24 Hour HelpLine for moreinformation.CareLink® Person

Página 182 - Index 167

You can share your reports with your healthcare professional. These reports can help you and yourhealthcare professional to discover trends and other

Página 183 - Index168

Troubleshooting and alarms/alertsAbout alarms and alertsTo receive the full benefits of your Guardian REAL-Time CGM System, all parts of the system mu

Página 184 - 6025254-018C_a

Follow these steps when you get an alarm or alert: To view the alarm/alert:1 From the HOME screen, review the alarm/alert by pressing any button to se

Comentários a estes Manuais

Sem comentários