Medtronic Guardian Real-time Guia do Utilizador Página 165

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 164
The following figure is a correlation plot of Guardian RT readings versus readings from the reference
method, the blood glucose meter. It is overlaid with the Clarke Error Grid. The total number of paired
data points is 2599.
Sensor BG (mg/dL)
A (1733, 66.7%)
B (665, 25.6%)
C (7, 0.3%)
D (193, 7.4%)
E (1, 0.0%)
Meter BG (mg/dL)
The percent of Guardian RT readings in the above graph are presented in the table below according to the
percentage of points falling within each zone (A-E). Results are further broken down (stratified) according
to the range of glucose concentrations.
Appendix A150
Vista de página 164
1 2 ... 160 161 162 163 164 165 166 167 168 169 170 ... 183 184

Comentários a estes Manuais

Sem comentários