Medtronic Guardian Real-time Guia do Utilizador Página 163

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 162
Agreement between the matched pair was estimated by evaluating the difference between the Guardian
RT reading and the blood glucose meter measurement. The difference between them was calculated as a
percentage of the blood glucose meter measurement (Mean Absolute Percent Difference). The bias was
also calculated, and it is defined as the overall difference between the Guardian RT glucose values and
the blood glucose meter values. The paired glucose measurements are summarized in the following table
Number of Paired Glucose Measurements 2599
Mean Absolute Percent Difference (± SD) 19.0 ± 19.7%
Bias -6.0 mg/dL (-0.3 mmol/L)
Appendix A148
Vista de página 162
1 2 ... 158 159 160 161 162 163 164 165 166 167 168 ... 183 184

Comentários a estes Manuais

Sem comentários