Warning: If you see any body fluid in the connector opening, you must discard the transmitter.
Because the transmitter contains a battery, do not discard in a bio-waste container. Instead,
continue to clean and disinfect the transmitter, and then discard according to local
regulations for battery disposal (non-incineration).
connector opening
connector pins
housing
3 Attach the tester to the transmitter.
4 If optional occlusive dressing was used and there is adhesive residue on the transmitter, follow the
instructions in Removing adhesive residue, on page 110.
5 Rinse the transmitter under running room temperature tap water for at least one minute and until any
visible debris is gone.
6 Prepare an enzymatic solution using two (2) ounces of ENZOL Enzymatic Detergent per gallon of
water.
7 With the tester still attached, fully submerge the transmitter in the enzymatic solution for one
minute.
System maintenance 107