Sensor glucose graph timeout
Default: 2 minutes
•Options
•2 minutes
•4 minutes
•6 minutes
NONE (no timeout)
High and low Glucose Limits (CSS7100)
Maximum number of Glucose Limits: 8
Low Glucose Limit
•Default
60 mg/dL (3.3 mmol/L)
•Range
40 to 390 mg/dL (2.2 to 21.6 mmol/L)
Increments
1 mg/dL (0.1 mmol/L)
High Glucose Limit
•Default
200 mg/dL (11.1 mmol/L)
•Range
50 to 400 mg/dL (2.8 to 22.2 mmol/L)
Increments
1 mg/dL (0.1 mmol/L)
High and low Glucose Limits (CSS7100K)
Maximum number of Glucose Limits: 8
Low Glucose Limit
•Default
90 mg/dL (5.0 mmol/L)
•Range
90 to 390 mg/dL (5.0 to 21.6 mmol/L)
Increments
1 mg/dL (0.1 mmol/L)
High Glucose Limit
•Default
280 mg/dL (15.6 mmol/L)
•Range
100 to 400 mg/dL (5.60 to 22.2 mmol/L)
Increments
1 mg/dL (0.1 mmol/L)
System specifications112