Medtronic Guardian Real-time Guia do Utilizador Página 119

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 118
5 Prepare a mild liquid soap solution using one (1) teaspoon of Ivory liquid soap per one (1) gallon of
room temperature tap water.
6 With tester still attached, submerge the transmitter in the mild liquid soap solution and soak for one
minute.
7 Holding the tester, brush the entire surface of the transmitter using a soft-bristled toddler toothbrush.
Make sure to brush all hard-to-reach areas until visibly clean.
8 Rinse the transmitter under running room temperature tap water for at least one minute, and until all
visible liquid soap is gone.
9 Dry the transmitter and tester with a clean, dry cloth.
System maintenance104
Vista de página 118
1 2 ... 114 115 116 117 118 119 120 121 122 123 124 ... 183 184

Comentários a estes Manuais

Sem comentários