Setting the BG Targets
The BG Target setting allows you to set
glucose targets. The Bolus Wizard will use
these targets to calculate a correction dose.
Because the targets may vary throughout the
day, your pump allows you to set up to eight
BG targets each day. If you want to set just
one target value, instead of a range, set both
the low and high values to the same number.
If your current BG is above the BG Target
range, the Bolus Wizard feature may
calculate a correction dose. A correction
dose will deliver enough insulin to bring your
BG down to your current high end of the BG
Target range. If your current BG is below the
BG Target range, the Bolus Wizard may
calculate a negative correction and subtract it from your food bolus. This will bring your BG to the low end
of the BG Target range.
Pumps are sent from the factory with default BG target range of 100-100 mg/dL (5.6-5.6 mmol/L).
Glucose Concentration
22.2
13.9
5.6
3.8
400
250
100
70
mmol/L mg/dL
Correct down to the upper
limit of 100 mg/dL
Correct up to the lower limit
of 70 mgl/dL
Target Range
1
Make sure the EDIT SETTINGS screen is open.
Main > Bolus > Bolus Wizard Setup > Edit Settings
2
Select BG Target, then press ACT.
3
The TARGET RANGE 1 screen appears. The low end of your BG Target range flashes on the screen.
4
Set the BG Target, then press ACT.
5
The high end of your BG Target range flashes on the screen. Enter the BG Target, then press ACT.
The start time for your first BG Target is midnight and cannot be changed.
6
a. If you adjust your BG Targets outside of 90-140 mg/dL (5.0-7.8 mmol/L), the pump screen displays
a warning that the values are acceptable but outside normal range. Press ESC to change your BG
target or press ACT to set this range.
b. If your BG target is within 90-140 mg/dL (5.0-7.8 mmol/L), the SET START TIME 2 screen appears.
The dashes under the screen name flash. The first BG target range is now set.
If you do not need a second BG Target range, press ESC and skip to the next section. If you need to
set up another BG Target range, follow steps 7 through 11.