7
In the SET START TIME 2 screen enter the time of the day you want this BG Target range to become
active.
8
Press ACT. The TARGET RANGE 2 screen appears.
9
The low end of your BG Target range flashes on the screen. Set the BG Target, then press ACT.
10
The high end of your BG Target range flashes on the screen. Enter the BG Target, then press ACT.
11
The SET START TIME 3 screen appears. The second BG Target range is now set.
12
If you do not need to set up any more BG Target ranges, press ESC. If you need to set up more BG Target
ranges, repeat steps 7 through 11 above for each one.
You are now ready to set up the Active insulin time.
About active insulin
Active insulin is the bolus insulin that has already been delivered to your body, but has not yet been used.
The pump considers your active insulin time setting in determining any active insulin still in your body from
prior boluses. This may help prevent hypoglycemia caused by over-correcting for high blood glucose.
The Bolus Wizard feature automatically tracks active insulin for you, based on your active insulin time, and
subtracts the appropriate amount when your blood glucose is above your target range. The details will appear
in the ESTIMATE DETAILS screen during the bolus programming steps.
Your Paradigm pump is shipped from the factory with an active insulin time setting of six hours, which most
closely matches the published scientific data. If your healthcare professional prescribes a different time for
you, the active insulin time setting can be adjusted in the Bolus Wizard menu in one-hour increments from
two to eight hours.
For more details about active insulin, see the Bolus Wizard feature specifications section in the Pump
specifications chapter.
CAUTION: If you give yourself insulin by using a syringe, the Bolus Wizard feature will not be able to
correctly determine the active insulin in your system. Consult with your healthcare professional on how
long you need to wait after a manual injection before you can rely on the active insulin calculation of
your Bolus Wizard feature.
Using the Bolus Wizard feature