Entering meter BG
Sensor calibration will be successful only if the BG entry is in the range of 40-400 mg/dL (or 2.2 to 22 mmol/L).
You should calibrate three to four times spread throughout the day for optimal results.
1
ಘ
Use one of these ways for all the calibrations throughout the sensor's life, including the first calibration
after the initialization.
a. To enter a meter BG value manually, follow this path:
Main > Sensor > Enter Meter BG
b. If at the time of sensor calibration you also need to perform a bolus using the Bolus Wizard, you can
use the meter BG for both calibrating the system and bolusing. To enter a meter BG value manually
in the Bolus Wizard, press the
button on your pump, or use the MAIN MENU to go to the ENTER BG
screen:
Main > Bolus > Use Bolus Wizard
c. To enter a BG value automatically through the linked meter, test your BG with a fingerstick. The
value will be automatically sent from the meter to your pump. Make sure that your pump and the
meter are properly programmed for communication. See the Meter option section in the Using the
Bolus Wizard feature chapter for more information.
2
ಘ
If you are manually entering your meter BG, use the up and down arrow buttons on your pump to do so.
Press ACT. For the Bolus Wizard option above, select Yes when the following screen displays: BG TO
UPDATE SENSOR
Sensor features
127
ಘChapter 8