Medtronic Paradigm Link Guia do Utilizador Página 182

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 181
3
If the condition is due to a low battery, replace the battery.
4
If the condition is due to a low reservoir, monitor the reservoir volume frequently and change the
reservoir when appropriate. Make sure you have a new reservoir, infusion set and vial of insulin with you.
Pump alert conditions
Listed below are alerts that indicate a condition that is outside normal pump activity.
LOW RESERVOIR
You can program the pump to sound an alert when either a specified number of units remains or a specified
amount of time remains before the reservoir will be empty.
LOW BATTERY
If you get this alert, do not go to sleep without replacing the battery. The backlight, the remote, and meter
functions are disabled during a LOW BATTERY condition. If the alert type is set to Vibrate, the pump will
change to the audio alert Beep-medium. Clear (ESC, ACT) this alert before you replace your battery.
Sensor alert conditions
Listed below are the alerts that you may encounter while using the sensor feature of your pump, and how to
resolve the alert condition.
WEAK SIGNAL
Reason: Occurs when the pump does not receive data from the transmitter for a pre-defined period of time
(as set in Missed Data).
Action: Move the pump closer to the transmitter or move the transmitter and the pump to the same side of
your body.
Troubleshooting and alarms
163
Chapter 13
Vista de página 181
1 2 ... 177 178 179 180 181 182 183 184 185 186 187 ... 261 262

Comentários a estes Manuais

Sem comentários