Example
You set your High Snooze to 15 minutes. You receive an alert at 1:00 pm and immediately clear it. However,
the high glucose condition that caused the alert still exists, so you will receive a second alert at 1:15 pm, 15
minutes after you cleared the first alert. If you immediately clear the alert again, then the alert will repeat
at 1:30 pm and continue to repeat every 15 minutes until you resolve the condition that caused the alert.
(This example also applies to the Low Snooze.)
ಗ
To set up your High Snooze:
1
ಘ
Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Sensor Setup > Edit Settings
2
ಘ
Select High Snooze, then press ACT. The SET HIGH SNOOZE screen flashes the default High Snooze time
of 1 hour (1:00).
3
ಘ
Select your High Snooze time. The time must be between five minutes (0:05) and three hours (3:00).
4
ಘ
Press ACT. The EDIT SETTINGS screen shows the High Snooze time set.
You are now ready to set up your Low Glucose limit.
Low Glucose Alert
Your Low Glucose alert must be turned on if you want your pump to alert when the sensor glucose
measurement reaches or goes below your set Low Glucose limit.
ಗ
To turn on your Low Glucose alert, and to set the Low Glucose limit:
1
ಘ
Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Sensor Setup > Edit Settings
2
ಘ
Select Low Glucose, then press ACT.
3
ಘ
Select On, then press ACT. The SET LOW GLUCOSE screen flashes the default Low Glucose limit of 50
mg/dL (2.8 mmol/L).
4
ಘ
Select your Low Glucose limit between 40–390 mg/dL (2.2–21.6 mmol/L). Your Low Glucose limit must
be at least 10 mg/dL (0.6 mmol/L) below your High Glucose limit.
5
ಘ
Press ACT. The EDIT SETTINGS screen shows the Low Glucose value set. You are now ready to set up your
Low Snooze.