Continuous Glucose Recorder User Guide 9
Charger
WARNING: Always clean the digital recorder after removing it from the patient
and before attaching it to the charger. See the cleaning instructions on page 20.
The digital recorder contains a non-replaceable, rechargeable battery that you can
recharge as needed with the charger. The charger has a green light that shows the
charging status and a red light that indicates any problems during charging. If you see a
red light, see the Troubleshooting section on page 18. The charger needs one AAA
battery, size E92, type LR03, to operate. A new AAA battery contains enough power to
recharge the digital recorder more than 40 times.
Installing a new charger battery
1. Remove the charger battery cap by turning it
counter-clockwise 1/4 turn using a coin in the
groove of the cap.
2. Insert a new AAA battery with the flat (-) end
first. Make sure that you align the small bumps
on the battery cap with the small notches in the
charger’s battery opening. Push in the cap all
the way using a coin. Turn the cap clockwise 1/4
turn to close.
3. If the battery is installed incorrectly or is low, the charger will not work. Repeat the
steps above using a new battery.
Charging the digital recorder
Before using the digital recorder for the first time, you must fully charge the digital
recorder battery which may take up to eight hours. It is recommended to recharge the
digital recorder after each glucose sensor use. If you choose to recharge the digital
recorder after a 3-day glucose sensor use, the charging time will be less than 20
minutes.