10 CGMS
®
iPro
TM
Continuous Glucose Recorder User Guide
1. If a green light on the digital recorder is lit or flashing, do not connect it to the charger.
The digital recorder will not charge with its green light on. Wait for the green light to
turn off (approximately 30 seconds), then connect the digital recorder to the charger.
2. Connect the digital recorder to the
charger by lining it up, flat side down,
with the charger. Push the two
components together fully. Always
allow at least one minute before
disconnecting the digital recorder
from the charger or the digital
recorder may not work properly. If you
disconnect the digital recorder before
one minute, reconnect it to the
charger for at least one minute.
3. Within 10 seconds after the digital
recorder is connected, a green light on
the charger will flash for 1-2 seconds as
the charger powers on. For the rest of
the charging time, the charger’s green
light will flash in a continuing pattern of
four flashes, pause, four flashes, pause.
4. When charging is complete, the green
light on the charger will stay on, without
flashing, for 15-20 seconds and then turn off.
5. After the green charger light turns off,
disconnect the digital recorder from the
charger. The green light on the digital recorder
will flash for about five seconds, then turn off.
6. If the green light on the digital recorder does not
flash, reconnect it to the charger for at least one
minute.
7. After removing the digital recorder from the charger, wait at least one minute before
connecting it to a sensor or tester.
Green light
Green light