Medtronic iPro2 Guia do Utilizador Página 17

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
Vista de página 16
CGMS
®
iPro
TM
Continuous Glucose Recorder User Guide 15
5. When connecting the digital recorder to the glucose sensor, touch the inserted glucose
sensor at the back of the assembly to prevent movement.
6. Hold the digital recorder as shown. Line up the
two notches on both sides with the flexible side
arms of the glucose sensor. The digital
recorder flat side with the label should face
the skin.
7. Slide the digital recorder onto the glucose
sensor. Push it in firmly until the flexible side
arms of the glucose sensor snap into the
notches on both sides of the digital recorder. In
the next 20 seconds the digital recorder will
flash green for about 10 seconds. This indicates
that the digital recorder is properly connected to a sensor and that it has enough
battery power to work for at least three days.
8. If the digital recorder light does not flash, disconnect it from the glucose sensor. Wait
for at least one minute and then reconnect. If the digital recorder light still does not
flash, charge the digital recorder.
9. After the digital recorder light flashes green, you
will be prompted by Solutions iPro to swipe the wand
across the digital recorder. Please refer to Using the
CGMS iPro Recorder and iPro Wand, on page 13, for
correct swiping technique.
10. Solutions CGMS iPro will tell you if you have
successfully started communication or if you need to
repeat the start process.
11. After the digital recorder is successfully initialized
by Solutions CGMS iPro, the Solutions software will
prompt you to apply the recommended occlusive dressing over the digital recorder and
the glucose sensor.
Sensor