14 CGMS
®
iPro
TM
Continuous Glucose Recorder User Guide
Communicating with Solutions CGMS iPro
You u se So lut io ns CGMS iPro to set up a new patient and to download data from the
digital recorder. You should always refer the Solutions Software for CGMS iPro
Continuous Glucose Recorder User Guide for complete instructions. The digital recorder
and Solutions CGMS iPro must communicate during initialization and data transfer. This
section provides an overview of how the components are used for each process.
CAUTION:This procedure describes how to connect the glucose sensor to the digital
recorder. It is very important that you follow the sequence provided by Solutions
CGMS iPro when setting up the device. Solutions CGMS iPro will indicate when to
connect the devices. See your Solutions Software for CGMS iPro Continuous Glucose
Recorder User Guide for more details.
Setting up the digital recorder
1. Enter the digital recorder serial number into the
Solutions software. The serial number is on the flat
side of the digital recorder, as shown on the right.
Enter only the seven numbers, not the letters, when
setting up a new patient.
Connecting the glucose sensor
CAUTION: Do not connect the digital recorder to the
glucose sensor until you are prompted to do so by
Solutions CGMS iPro.
1. It is recommended that the digital recorder be fully charged before connecting it to the
glucose sensor. If it is not fully charged, it may not flash when connected to the sensor
and may not operate the sensor for the entire three days.
2. First insert the glucose sensor (follow the instructions provided with the glucose
sensor). Check for bleeding. Make sure that bleeding has stopped (see Warnings on
page 5 for details). Do not connect the digital recorder at this time.
3. The glucose sensor must be inserted in the patient’s body for at least five minutes
before connecting the digital recorder.
4. When prompted by the Solutions software, connect the digital recorder to a glucose
sensor already inserted into the body. The digital recorder will not work if connected to
a dry sensor (not inserted).