16 CGMS
®
iPro
TM
Continuous Glucose Recorder User Guide
Downloading data from the digital recorder
Follow the Solutions software instructions to download the patient’s data from the
digital recorder.
1. Setup the COMLINK iPro. Make sure that the COMLINK iPro is connected to your PC or
the optional serial switch box.
2. If you are using the serial switch box, make sure that you select
the communications port that corresponds to the COMLINK iPro.
If the digital recorder has been disconnected from the sensor,
then connect it to the blue tester. Whichever way you choose to
download, make sure that the digital recorder is as close as
possible to the COMLINK iPro to ensure a good RF connection.
Click NEXT to continue.
3. When the software prompts you, swipe the digital recorder with
the iPro Wand to start communication and begin the download.
4. After a successful data download, clean and dry the digital recorder, remove the tester,
and connect the digital recorder to the charger.
Calibrating Solutions CGMS iPro
The Solutions software must be calibrated with fingerstick measurements to maximize
sensor reading accuracy. The Solutions software requires at least four fingerstick
measurements per 24-hour period for each of the three-day study interval (a minimum
total of 12 fingerstick measurements).
Be sure patients know to take fingerstick glucose measurements when their blood
glucose is stable - such as before meals and at bedtime, and to avoid times such as right
after eating, exercising or insulin delivery.
The Solutions software only accepts BG meter values between 2.2-22 mmol/L (40 and
400 mg/dL).
If patients use one of the Solutions-compatible BG meters listed in this user guide, the
BG measurements will automatically download from the meter into the Solutions
software during the download process.
Patients who use a non-compatible BG meter must log each BG meter value in the BG
Meter and/or in their Patient Log sheet as soon as they take a reading. This allows the
Comentários a estes Manuais