Continuous Glucose Recorder User Guide 23
Guidance and Manufacturers Declaration – Electromagnetic Immunity
The CGMS iPro Digital Recorder (MMT-7709) is intended for use in the electromagnetic environment
specified below. The customer or the user of the CGMS iPro Digital Recorder should assure that it is
used in such an environment.
Immunity Test
IEC 60601 Test
Level
Compliance
Level
Electromagnetic Environment
Guidance
Electrostatic
discharge (ESD)
±6 kV contact Not applicable The CGMS iPro Digital Recorder should
not be affected by electrostatic
discharge that might occur under normal
conditions of use.
IEC 61000-4-2 ±8 kV air ±30 kV air
(<5% relative
humidity)
Electrical fast
transient/burst
±2 kV for power
supply lines
Not applicable
IEC 61000-4-4 ±1 kV for input/
output lines
Not applicable
Surge ±1 kV line(s) to
line(s)
Not applicable
IEC 61000-4-5 ±2 kV line(s) to
earth
Not applicable
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
Not applicable
40% U
T
(60% dip in U
T
)
for 5 cycles
Not applicable
70% U
T
(30% dip in U
T
)
for 25 cycle
Not applicable
<5% U
T
(>95% dip in U
T
)
for 5 seconds
Not applicable
Power frequency
(50/60 Hz) magnetic
field
IEC 61000-4-8
3 A/m 3 A/m Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
NOTE:
U
T
is the a.c. mains voltage prior to application of the test level.