Guidance and Manufacturers Declaration – Electromagnetic Immunity
The CGMS iPro Digital Recorder (MMT-7709) is intended for use in the electromagnetic environment
specified below. The customer or user of the CGMS iPro Digital Recorder should assure that it is used in
such an environment.
Immunity
Test
IEC 60601
Level
Compliance
Level
Electromagnetic Environment Guidance
Conducted RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3Vrms
150kHz to
80MHz
3 V/m
80MHz to 6GHz
Not
applicable
3 V/m
Portable and mobile RF communications equipment
should be used no closer to any part of the CGMS iPro
Digital Recorder, including cables, than the
recommended separation distance calculated from
the equation applicable to the frequency of the
transmitter.
Recommended separation distance:
80 MHz to 800 MHz
800 MHz to 6 GHz
Where P is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer and d is the recommended separation
distance in meters (m).
Field strengths from fixed RF transmitters, as
determined by an electromagnetic site survey
a
,
should be less than the compliance level in each
frequency range.
Interference may occur in the vicinity of equipment
marked with the following symbol:
L
a. Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones
and land mobile radios, amateur radio, AM and FM radio broadcasts and TV broadcast cannot be
predicted theoretically with accuracy. To access the electromagnetic environment due to fixed RF
transmitters, an electromagnetic site survey should be considered. If the measured field strength in the
location in which the CGMS iPro Digital Recorder is used exceeds the applicable RF compliance level
above, the CGMS iPro Digital Recorder should be observed to verify normal operation. If abnormal
performance is observed, additional measures may be necessary, such as re-orienting or relocating the
CGMS iPro Digital Recorder.
NOTE:
At 80 MHz and 800 MHz, the higher frequency range applies.
NOTE: These guidelines may not apply in all situations. Electromagnetic propagation is
affected by absorption, and reflection from structures, objects and people.
d 1.2 P=
d 2.3 P=