26 CGMS
®
iPro
TM
Continuous Glucose Recorder User Guide
Specifications
Biocompatibility
Digital Recorder: Complies with ISO 10993-1 for body contact
Operating
Conditions
Digital Recorder Temperature: +32° to +122°F (0° to 50°C)
Digital Recorder Relative Humidity: 10% to 95% with no condensation
Charger Temperature: +50° to +104°F (10° to 40°C)
Charger Relative Humidity: 30% to 75% with no condensation
Storage
Conditions
Digital Recorder Temperature: -4° to +131°F (-20° to +55°C)
Digital Recorder Relative Humidity: 10% to 100% with condensation
Charger Temperature: +14° to +122°F (-10° to +50°C)
Charger Relative Humidity: 10% to 95% with no condensation
Battery Life
Digital Recorder: 14 days of continuous glucose monitoring
immediately following a full charge
Charger: Completes 40 typical charging operations using a new AAA
battery
Digital Recorder
Frequency
MMT-7709NA - 916.5 MHz
MMT-7709WW - 868.35MHz
Digital Recorder
Dimensions and
Weight
Width: 1.40 inches (3.56 centimeters)
Length: 1.12 inches (2.84 centimeters)
Height: 0.37 inches (0.94 centimeters)
Weight: 0.2 ounces (5.67 grams)
ComLink iPro
Frequency
MMT-7304NPRO - 916.5 MHz
MMT-7304WPRO - 868.35 MHz
Comentários a estes Manuais