20 CGMS
®
iPro
TM
Continuous Glucose Recorder User Guide
Question: Why didn’t I see the digital recorder green light flash after connecting it to
the blue tester?
Answer: Check the connection. If you still do not see a green light flash, fully recharge
the digital recorder battery. Test the digital recorder with the tester, as described on
page 17 (make sure the tester is clean before using). If you still do not see a green light
flash, clean the tester connector. If you still do not see a green light flash, call the 24
Hour HelpLine or your local representative. It may be time to replace your digital
recorder.
Cleaning the CGMS iPro Digital Recorder
The CGMS iPro Digital Recorder is intended for multiple patient use. Follow this
procedure to clean the digital recorder after each patient use.
CAUTIONS:
The charger and the tester are NOT watertight. Do NOT immerse in water.
Do NOT discard the digital recorder in a medical waste container or otherwise
subject it to incineration. The digital recorder contains a battery which may explode
upon incineration.
1. Attach the tester to the digital recorder to make sure that NEITHER water NOR
disinfectant NOR alcohol gets inside the digital recorder’s connector.
2. Dampen a clean cloth with a mild liquid soap solution. Wipe the outside of the digital
recorder.
3. Rinse the digital recorder under warm tap water but do NOT get water inside the
connector. If you get water inside the connector, shake the water out and allow it to air
dry.
4. Apply three to four drops of a quaternary ammonium compound disinfectant (for
example, Cavicide
®
) on a clean, dry cloth and wipe the digital recorder.
5. Hold the digital recorder and rinse with 70% isopropyl alcohol.
6. If there is adhesive residue on the digital recorder, you can remove it with Detachol
®
adhesive remover between each patient use.
7. Disconnect the tester from the digital recorder.
8. Place the digital recorder on a clean, dry, non-shedding cloth and air dry.