Medtronic Revel Guia do Utilizador Página 104

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 265
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 103
4 For a Square Wave bolus, do these steps:
a. Select Square Wave Bolus, then press ACT. The SET SQUARE BOLUS screen appears.
b. Enter the desired amount for the Square Wave bolus units, then press ACT.
c. Continue to step 5.
For a Dual Wave bolus, do these steps:
a. Select Dual Wave Bolus, then press ACT. The SET DUAL BOLUS TOTAL screen appears.
b. Enter the desired amount for the total dual bolus units. This amount is the total of Normal and
Square Wave bolus units. Press ACT.
c. The next screen flashes the amounts of Now (Normal) and Square Wave portions of the Dual Wave
bolus. The screen also shows the percentage amount of each portion. Press
or to change
the percentage/number of units. Continue to step 5.
5
The SQUARE DURATION screen appears. Enter the amount of time you want the Square Wave bolus to
last, then press ACT.
NOTE: If you have BG Reminder turned On, the BG REMINDER DURATION screen displays. It allows
you to set the duration before you are reminded to check your blood glucose after a bolus. See the
BG Reminder section in the Basic programming chapter for information about this feature.
6 The BOLUS DELIVERY screen appears with an open circle indicating that your pump is in Special mode.
The pump beeps/vibrates at the start of the bolus. During bolus delivery, the pump will return to the
HOME screen. The pump beeps/vibrates at the end of the bolus and the open circle disappears.
Optimizing pump therapy 83
Chapter 6
Vista de página 103
1 2 ... 99 100 101 102 103 104 105 106 107 108 109 ... 264 265

Comentários a estes Manuais

Sem comentários