
WEAK BATTERY
The pump tests the voltage of each battery installed. If the battery voltage is less than full strength, this
alarm may occur. The pump will operate normally, but the battery life will be shorter than expected.
Always make sure to install a new battery in the pump.
Testing the transmitter
The tester acts like a sensor. If you get sensor-related alerts, use the tester to make sure that the
transmitter is working.
Connecting the tester
1 Hold the transmitter and the tester as shown. Line up the flat side of the tester with the flat side of
the transmitter.
2 Push the tester into the transmitter until the flexible side arms of the tester click into the notches on
both sides of the transmitter.
3 Within 20 seconds, the green light on the transmitter will flash for about 10 seconds when properly
connected. The Sensor feature must be turned On on your pump. Go to the RECONNECT OLD SENSOR
screen to start your sensor.
Main > Sensor > Link to Sensor > Reconnect Old Sensor
You should see the sensor icon
on the pump screen, which means that the transmitter and the
sensor are communicating.
Troubleshooting, alerts and alarms 169
Chapter 13
Comentários a estes Manuais