4 Within the next seven to eight minutes, go to the SENSOR STATUS screen to look for the Sensor ISIG
value:
a. The sensor ISIG value on this screen should be between 24.00–29.00 nA. This range of ISIG values
means that the transmitter electronics are working properly. Since the transmitter is sending
signals correctly, the sensor must have caused the alert. Remove and discard the sensor. Insert a
new sensor in a new site.
b. If you see a Sensor ISIG value that is less than 24.00 nA or more than 29.00 nA, contact our
24 Hour HelpLine. It may be time to replace your transmitter.
Disconnecting the tester
1 Hold the transmitter body as shown and pinch the side arms of the tester.
2 With the tester arms pinched, gently pull the transmitter away from the tester. To save transmitter
battery life, do not leave the tester connected after testing.
Troubleshooting sensor features
Reconnect old sensor
You should only use this feature if you have disconnected the sensor from the transmitter and have to
reconnect them.
To Reconnect old sensor:
1 Go to the RECONNECT OLD SENSOR screen.
Main > Sensor > Link to Sensor > Reconnect Old Sensor
Troubleshooting, alerts and alarms170