Guidance and Manufacturer's Declaration - Electromagnetic Immunity
The Paradigm 523/723/523K/723K pump system (consisting of the MMT-523/723/523K/723K pump
and MMT-7703 transmitter) is intended for use in the electromagnetic environment specified below.
The customer or the user of the Paradigm 523/723/523K/723K pump system should make sure that
it is used in such an environment.
Immunity Test IEC 60601 Test Level Compliance
Level
Electromagnetic Environment
- Guidance
Electrostatic discharge
(ESD)
IEC 61000-4-2
±6 kV contact
±8 kV air
±30 kV air
(<5% relative
humidity)
The Paradigm
523/723/523K/723K pump
system should not be affected
by electrostatic discharge that
might occur under normal
conditions of use.
Electrical fast transient/
burst
±2 kV for power supply
lines
Not applicable Not applicable
IEC 61000-4-4 ±1 kV for input/output
lines
Not applicable Not applicable
Surge ±1 kV line(s) to line(s) Not applicable Not applicable
IEC 61000-4-5 ±2 kV line(s) to earth Not applicable Not applicable
Voltage dips, short
interruptions and
voltage variations on
power supply lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
)
for 0.5 cycle
Not applicable Not applicable
40% U
T
(60% dip in U
T
)
for 5 cycles
Not applicable Not applicable
70% U
T
(30% dip in U
T
)
for 25 cycles
Not applicable Not applicable
<5% U
T
(>95% dip in U
T
)
for 5 seconds
Not applicable Not applicable
Pump specifications 199
Chapter 15