HIGH XXX MG/DL (XXX = SG measurement)
Reason: The sensor glucose number is higher than or equal to your high glucose limit. If you do not set a
High Glucose Limit and do not set the glucose alerts ON, then you will not get a High sensor glucose alert.
The pump plays four consecutive tones, in rising pitch, if an audible beep has been selected as the alert
type.
LOW PREDICTED
Reason: The sensor glucose measurements may reach or go below your Low Glucose Limit in the length of
time you selected for the Low Predictive alert. The pump plays three consecutive tones, in falling pitch, if
an audible beep has been selected as the alert type.
HIGH PREDICTED
Reason: The sensor glucose measurements may reach or go above your High Glucose Limit in the length of
time you selected for the High Predictive alert. The pump plays three consecutive tones, in rising pitch, if
an audible beep has been selected as the alert type.
RISE RATE
Reason: The sensor glucose measurements are rising at a rate that is equal to or faster than the Set Rise
Rate Limit you selected for the alert. The pump plays two consecutive tones, in rising pitch, if an audible
beep has been selected as the alert type.
FALL RATE
Reason: The sensor glucose measurements are falling at a rate that is equal to or faster than the Set Fall
Rate Limit you selected for the alert. The pump plays two consecutive tones, in falling pitch, if an audible
beep has been selected as the alert type.