Setting up the reservoir
Warning: Do not use the reservoir or infusion set if any liquid gets on the top of the reservoir or inside
the tubing connector (as shown in the image). Liquid can temporarily block the vents. This may result
in the delivery of too little or too much insulin, which can cause hypoglycemia or hyperglycemia. If
any liquid gets on the top of the reservoir or inside the tubing connector, start over with a new
reservoir and infusion set.
Tubing
connector
Top of
reservoir
WARNING: Using cold insulin can cause air bubbles in the reservoir and tubing. If your insulin is stored
in the refrigerator, allow it to reach room temperature before filling your reservoir. When filling the
reservoir, take care to remove air bubbles.
Filling the reservoir
1 Remove the reservoir from the package. Make sure the plunger rod is fully extended.
2 Swab the vial with alcohol (not shown).
3 Making sure you do not push down on the plunger, press the transfer guard onto the vial.
plunger
rod
transfer
guard
reservoir
vial
31
4 Push down on the plunger to pressurize the vial. Hold down the plunger rod.
Starting on insulin52