1.50 U/H
(rate 1)
2.55 U/H
(rate 2)
1.25
U/H
1.95 U/H
(rate 3)
(50% x rate 3 = 0.975)
0.975 rounded down to
the next 0.05 increment
is 0.95 U/H
(50% x rate 2 = 1.275)
1.275 rounded down to
the next 0.05 increment is
1.25 U/H
7:00 a.m.
(07:00)
1:05 p.m.
(13:05)
3:00 p.m.
(15:00)
5:05 p.m.
(17:05)
temp basal stop time
12:00 a.m.
(00:00)
Temp Basal Settings
Temp basal type: Percent of basal
Duration: 4 hours (1:05 p.m. - 5:05 p.m.)
Rate: 50%
temp basal start time
0.95
U/H
Selecting temp basal type
Your pump will remember the temp basal type setting. Once the type is set, you do not have to set it
again. To select a temp basal type, do these steps:
1 Go to the SET TEMP BASAL AS screen.
Main > Basal > Basal Setup > Temp Basal Type
2 The SET TEMP BASAL AS screen appears. Select Insulin Rate (U/H) or Percent of Basal, then press
ACT.
Optimizing pump therapy98