SENSOR ERROR
Reason: The sensor signals are either too high or too low.
Action: You do not need to change the sensor. Clear the alert. If the alert persists, test your transmitter
with the tester.
Alarms
Your pump has a sophisticated network of safety checks and systems. If the safety network detects
anything unusual, your pump notifies you of conditions that require your immediate attention. The
backlight illuminates the pump screen and the alarm message displays on the screen.
It is important that you understand how your pump behaves when you do not clear an alarm or an alert.
See the Alerts section in this chapter for an explanation and an example.
Check your pump and test your blood sugar. Follow the guidelines prescribed by your healthcare
professional to manage your diabetes as needed.
NOTE: The STATUS screen shows any alarms and alerts that are active.
Question: Why are alarms important?
Answer: Your pump monitors activities and notifies you if there is an unusual pump status or your
attention is required. When an alarm is active, INSULIN DELIVERY IS STOPPED and immediate operator
interaction is required.
If the vibrate mode is on, all alarms and alerts start as vibrations and
then change to beeps. For your safety, if there is no response after
10 minutes, the beeps change to a siren. The siren continues every
minute until the alarm is cleared.
When a solid circle appears, follow
the instructions on the screen
Troubleshooting, alerts and alarms164