LOW TRANSMTR
Reason: Occurs when the transmitter battery is close to running out of power. This alert will repeat daily
at noon while this condition exists. The transmitter will continue sending sensor signals for several hours
and may last for several days until the battery becomes depleted. Be prepared to recharge your
transmitter immediately when its battery becomes depleted.
Action: Recharge your transmitter as soon as possible. See the Charging the transmitter section in the
Sensor features chapter.
CHARGE TRANSMTR
Reason: The transmitter battery is depleted.
Action: Recharge your transmitter immediately. See the Charging the transmitter section in the Sensor
features chapter.
CHANGE SENSOR
Reason: You may receive this alert after you receive two Cal Errors in a row, without the Cal Errors, or
when initializing your sensor.
Action: If the alert resulted from two Cal Errors in a row, replace the sensor. If the alert happened
without two Cal Errors, use the transmitter's tester to make sure your transmitter is working properly. If
you received this alert during initialization, you may be able to resolve this alert without replacing your
sensor. Call our 24 Hour HelpLine for further assistance.
SENSOR END
Reason: The sensor has reached the end of its life.
Action: Replace your sensor. The sensor has a maximum life of 72 hours (3 days). The 72-hour life span of
the sensor begins when the pump receives the first METER BG NOW alert.
Troubleshooting, alerts and alarms 161