•Range:
mg/dL/min
mmol/L/min
from 1.1 to 5.0
from 0.065 to 0.275
• Increments:
mg/dL/min
mmol/L/min
0.1
0.005
4 Press ACT. The SET RISE RATE LIMIT screen flashes the rate alert default setting: OFF.
5 Select the Rise Rate Limit. The range and increments are the same as the Fall Rate Limit (see step 3
above).
6
Press ACT. The Rate of Change alerts will now use the settings you specified. You are now ready to set
up the Cal Repeat.
Setting the Cal Repeat
After you receive and clear a METER BG NOW alert, the device will repeat the alert until you enter a new
meter blood glucose measurement. The Cal Repeat feature allows you to set how frequently you want the
alert to repeat after you clear it.
1 Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Cal Repeat, then press ACT. The SET CAL REPEAT screen flashes the default Cal Repeat time of
30 minutes (0:30).
3 Set your Cal Repeat time between 5 minutes (0:05) and 1 hour (1:00).
4 Press ACT. The EDIT SETTINGS screen shows the Cal Repeat time you selected. You are now ready to
set your Cal Reminder.
Setting the Cal Reminder
The Cal Reminder feature allows you to set a reminder to calibrate your system. For example, if you set
your reminder to four hours, then you will receive a METER BG BY alert (Cal Reminder) four hours before
the next meter blood glucose (BG) entry is due (eight hours after your last successful sensor calibration).
To set up the Cal Reminder:
1
Make sure the EDIT SETTINGS screen is open.
Main > Sensor > Edit Settings
2 Select Cal Reminder, then press ACT. The CAL REMINDER screen shows On.
3 Press ACT. The SET CAL REMINDER screen flashes the Cal Reminder default time of 1 hour (1:00).
Sensor features110